Specification
*measure range: 0.00pH-14.0 pH
*resolution: 0.01ph
*accuracy: +/-0.01ph
*Power supply: 2*1.5V ( keep using more than 500 hours)
*operating temperature: 0℃--80℃
*calibration: two points automatic calibration
*dimension: 155mm*31mm*18mm
*weight: 50g
Operation
1.remove the protective cap.
2.first rinse the electrode with distilled water, and suck it with filter paper.
3.turn the meter on by pressing the “ON” key.
4.immerse the ph meter electrode in the solution to be tested(can not be over the immersion line )
5.stir gently and wait around 30 seconds till the reading stabilized.
6.after finished ,clear the electrode with pure water ,turn the meter off by pressing the “ON/OFF” key.
7.Always replace the protective cap after use.
Calibration
1. Turn on pH meter
2. Disolve each buffer solution in 250 ml of distilled water .
3. Immerse the electrode into the pH 6.86 solution.
(under the temperature of 25)
4. Press the "CAL" (calibration) button for 3 seconds and release
-Display will start flashing 6.86
-Wait until the display stops flashing.
Rinse the electrode with distilled water and dry it with filter paper.
5. Immerse the electrode in pH 4.00 solution,
- Press "CAL" button for 3 seconds ,then press and release immediately second time
- Reading will start flashing 4.00
- Wait until display stops flashing
Rinse and dry the electrode with distilled water as before.
6. Test calibration by inserting electrode into pH 9.18 buffer solution,
gently swirl electrode until digital display stabilizes. Display should read between 9.16 and 9.20.
(if necessary repeat calibration procedure)
Attention
The instrument need recalibrate in the following condition.
1.Long time no use or use very long time after calibration
2.The electrode is used very frequently
3.The testing accuracy requirement is very high
4.Press the "CAL"button long time in the air carelessly
Maintenance:
Always replace protective cap after using digital meter to keep electrode from drying out due to prolonged exposure to air,
which leads to slow and/or unstable readings.
If electrode has been dried out immerse it into distilled water for a few hours.
LowBatt
When the display value is fuzzy or unshown , The battery should be replaced promptly.Pay attention to the polarity of battery.
Ако притежавате този продукт и имате впечатления за него, моля споделете ги!
Задайте своя въпрос относно PH Meter PH-02, Уред за измерване киселинност на вода
виж всички измервателни уреди
Тони Николов
Обслужването е на ниво,доставката е бърза ,цените са ниски в сравнение с по големи вериги.Препоръчвам
преди седмица
Silviya Nedelcheva
Бърза доставка и много бърза реакция при проблем. Много отзивчив персонал.
преди 4 месеца
Slav Ilchev
Бързо обработват и доставят поръчките, много професионално отношение.
преди 2 месеца
Dancho Danalev
Добро отношение. Намерих тоеа което търсех
преди 2 месеца
Georgi Petrov
Само хубави неща мога да кажа за тях! Бързо, коректно и на добра цена. Обслужването е на топ ниво! С удоволствие ще пазаря отново при Вас!
преди година